Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. FTMC C-SECTION TRAY-LF DYNJ54603B Recall
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. FTMC C-SECTION TRAY-LF DYNJ54603B
FDA recall Class II
Ongoing
FDA lists this Class II recall for Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. FTMC C-SECTION TRAY-LF DYNJ54603B. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2596-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 27, 2026
- FDA report date
- Jul 1, 2026
- Quantity in commerce
- 113, 843 kits
- Distribution
- See key details
- FDA cause
- No Marketing Application
What is being recalled?
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. FTMC C-SECTION TRAY-LF DYNJ54603B
Affected model & UDI checker
Search the exact identifiers published for Z-2596-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. FTMC C-SECTION TRAY-LF DYNJ54603B | Check source notice | 10195327458188 |
FDA code information
DYNJ54603B UDI-DI 10195327458188 Lot 24ABO124
Why was this product recalled?
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
What actions did Medline Industries, LP take?
This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2596-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology