Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542. Recall
Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2602-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 22, 2026
- FDA report date
- Jul 1, 2026
- Quantity in commerce
- 7 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542.
Affected model & UDI checker
Search the exact identifiers published for Z-2602-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 1000542 | Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542. | Check source notice | 813502013467 |
FDA code information
Impella Set Product Code: 1000542; GTIN: 813502013467; Serial numbers: N/A. Batch Number: 2041530. Introducer Product Code: 1000542; Batch Number: 2041530.
Why was this product recalled?
Potential for thrombus formation during prolonged use of the introducer.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Abiomed, Inc. take?
Abiomed notified consignees on about 05/22/2026, via letter titled "URGENT MEDICAL DEVICE RECALL (REMOVAL)." Consignees were made aware of the issue and that if thrombus formation were to occur, the user may encounter difficulty aspirating the sheath sidearm or valve and may observe thrombus in the syringe or sidearm post aspiration or around the valve post CP pump removal. They were instructed to review inventory for affected units and quarantine them, arrange for their return using the provided shipment label from Sedgwick, and complete and return the provided business response form. Consignees were also instructed to forward the notification to anyone in the facility that needs to be informed, if any of the subject products have been forwarded to another facility, contact that facility and provide them with the notification, and to post a copy of the notification in a visible area for awareness.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2602-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026