Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2609-2026

Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B Recall

Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2609-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 28, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
9,272 kits
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B

Affected model & UDI checker

Search the exact identifiers published for Z-2609-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936BCheck source notice10193489546606
Not specifiedMedline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936BCheck source notice10193489545678
Not specifiedMedline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936BCheck source notice10193489422818
FDA code information

DYNJ905855 UDI-DI 10193489546606 Lots 21ABU297 21BBB235 21DBV520 21FBG162 21GBC672 21GBI192 21IBT387 21VBB580 22DBK812 22FBR093 22FBS035 22HBI628 22IBV291 22JBX729 22NBE430 23ABD954 23BBI465 23CBO951 23CBX474 23GBF721 DYNJ905857 UDI-DI 10193489545678 Lots 21ABG510 21GBK940 DYNJ905936B UDI-DI 10193489422818 Lots 21BBF766 21DBP452 21FBF248 21LBJ459 22ABV319 22DBS751 22FBU868

View the original FDA recall notice

Why was this product recalled?

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

What actions did Medline Industries, LP take?

On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.