Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2610-2025

Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH Recall

Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH

FDA recall Class I

Ongoing

FDA lists this Class I recall for Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2610-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jul 31, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
7913 eaches
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Item Number/UDI (ea)/UDI(case): 401575RH 10888277407497 20888277407494; 401582RH 10888277407510 20888277407517; 401600RH 10888277407541 20888277407548; 401603RH 10888277407558 20888277407555; 401606RH 10888277407565 20888277407562; 401652RH 10888277407572 20888277407579; 401653RH 10888277407589 20888277407586; 401654RH 10888277407596 20888277407593; 401904RH 10888277407602 20888277407609; 401905RH 10888277407619 20888277407616; 401908RH 10197344020966 20197344020963; 401914RH 10888277407527 20888177407524; 401915RH 10888277407626 20888277407623; 401918RH 10888277407640 20888277407647; 401932RH 10888277407671 20888277407678; 401933RH 10888277407688 20888277407685; 401934RH 10888277407695 20888277407692; 401938RH 10888277407701 20888277407708; 401939RH 10888277407718 20888277407715; 401940RH 10888277407480 20888277407487; 401941RH 10888277407725 20888277407722; 401991RH 10888277407756 2088277407753; LOTS EP230724 EP230731 EP230803 EP230816 EP230914 EP231017 EP231116 EP240108 EP240122 EP240212 EP240227 EP240311 EP240322 EP240411 EP240415 EP240422 EP240429 EP240507 EP240514 EP240520 EP240528 EP240607 EP240611 EP240710 EP240715 EP240723 EP240805 EP240812 EP240820 EP240829 EP240911 EP240923 EP240930 EP241008 EP241030 EP241125 EP250108

View the original FDA recall notice

Why was this product recalled?

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

On July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.