FDA RECALL RECORD · Z-2615-2025

Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201 Recall

Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201 Model Number: 2001 UDI-DI code: 07350137521019 Inserter Adapter OTD Pinnacle Straight REF 30-202 Model Number: 1001 UDI-DI code: 07350137520593 Reamer Adapter OTD REF 30-157 Model Number: 1003 UDI-DI code: 07350137520081 Inserter Adapter OTD REF 30-158 Model Number: 1004 UDI-DI code: 07350137520098 The Inserter Adapters are intended to be attached to a Cup Inserter Tool. The Cup Inserter Tool (impactor) is intended to be tracked by and navigated to orient and position the hip implant cup correctly relative to preoperative planned position and orientation.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2615-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 28, 2025
FDA report date
Oct 1, 2025
Quantity in commerce
261 devices
Distribution
See key details
FDA cause
Process control

What is being recalled?

Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201 Model Number: 2001 UDI-DI code: 07350137521019 Inserter Adapter OTD Pinnacle Straight REF 30-202 Model Number: 1001 UDI-DI code: 07350137520593 Reamer Adapter OTD REF 30-157 Model Number: 1003 UDI-DI code: 07350137520081 Inserter Adapter OTD REF 30-158 Model Number: 1004 UDI-DI code: 07350137520098 The Inserter Adapters are intended to be attached to a Cup Inserter Tool. The Cup Inserter Tool (impactor) is intended to be tracked by and navigated to orient and position the hip implant cup correctly relative to preoperative planned position and orientation.

Affected model & UDI checker

Search the exact identifiers published for Z-2615-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
2001Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
30-201Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
1001Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
30-202Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
1003Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
30-157Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
1004Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
30-158Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source noticeNot specified
Not specifiedOrtoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source notice07350137521019
Not specifiedOrtoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source notice07350137520593
Not specifiedOrtoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source notice07350137520081
Not specifiedOrtoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201Check source notice07350137520098
FDA code information

Model Number: 2001 Catalog Number: 30-201 UDI-DI code: 07350137521019 Lot Numbers: 20244803 Model Number: 1001 Catalog Number: 30-202 UDI-DI code: 07350137520593 Lot Numbers: 20245103 20245104 20245109 Model Number: 1003 Catalog Number: 30-157 UDI-DI code: 07350137520081 Lot Numbers: 20242001 20242008 20245119 Model Number: 1004 Catalog Number: 30-158 UDI-DI code: 07350137520098 Only distributed O.U.S.

View the original FDA recall notice

Why was this product recalled?

Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.

What actions did Ortoma AB take?

On 07/28/2025, the firm sent via email an "URGENT: MEDICAL DEVICE FIELD CORRECTION" Letter to customers informing them that the Insert Adapter OTD Straight pin may partially or completely break. This malfunction, if not identified, may result in displaying incorrect information of the reaming or cup placement by OTS Hip Guide. The values displayed on the screen may not correctly represent the actual angle and position of the tool. The magnitude of the resulting error can vary and depends on the level of the weld breakage. Customers are instructed to: 1. Discontinue use and quarantine affected units until the instructions in this letter have been read, understood and acknowledged. Complete and return the attached form via email to [email protected]. 2. Adhere to the following instructions: "Visual inspection: Inspect the device prior, during, and after surgery, for damages to the weld, see section visual inspection. "In case of partial or complete weld breakage, either use a replacement adapter or revert to conventional surgery. -Refer to IFU for recalibration procedure. Recalibration would lead to a slight delay in navigated surgery, estimated less than 3 minutes. -The consequence of reverting to conventional surgery would be that the user would not have the benefits of a navigated surgery. "If product should fail, ensure that item is cleaned and sterilized according to instructions and file a formal complaint. Visual inspection Inspect all affected devices to ensure that there are no visible cracks or scratches and that the pin is fixed. (Inspect the weld at a distance of around 40 cm.) -Inspect the devices prior to, during, and after surgery, see Figure 2. -A weld passes visual inspection if the pin is not possible to move and there are no visible cracks in the weld (Figure 2-a). -A weld fails the visual inspection if it is possible to move the pin and/or if there are visible cracks or breakage. For questions contact Ortoma at [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.