FDA RECALL RECORD · Z-2618-2025

Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above); Recall

Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);

FDA recall Class II

Ongoing

FDA lists this Class II recall for Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2618-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 28, 2025
FDA report date
Oct 1, 2025
Quantity in commerce
156 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);

Affected model & UDI checker

Search the exact identifiers published for Z-2618-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedExtractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);Check source notice08714729790280
Not specifiedExtractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);Check source notice08714729790266
FDA code information

Instructions for Use: 51878205; Pre-Printed Ouch: 51368417; 1. UPN (REF): M00547030 (9-12mm Below); UDI-DI: 08714729790280; Lot Numbers: 36398057; 2. UPN (REF): (1) M00547000(9-12mm Above); UDI-DI: 08714729790266; Lot Numbers: 36426495;

View the original FDA recall notice

Why was this product recalled?

The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.

What actions did BOSTON SCIENTIFIC CORPORATION take?

On August 28, 2025 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Instructions: 1. Immediately cease further distribution or use of any remaining affected devices. Segregate the units in a secure place until they can be returned. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Return affected devices.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology