Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 Recall
Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
FDA recall Class II
Ongoing
FDA lists this Class II recall for Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2620-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 15, 2025
- FDA report date
- Oct 1, 2025
- Quantity in commerce
- 40
- Distribution
- See key details
- FDA cause
- Component change control
What is being recalled?
Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
Affected model & UDI checker
Search the exact identifiers published for Z-2620-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CH00000001 | Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 | Check source notice | Not specified |
| P60034 | Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 | Check source notice | Not specified |
| Not specified | Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 | Check source notice | 07613327296167 |
FDA code information
UDI-DI: 07613327296167 and 07613327651317 REF/Serial Numbers: CH00000001/7000010116, 7000010117, 7000010118, 7000010119, 7000010120, 7000010121, 7000010122, 7000010123, 7000010124, 7000010125, 7000010141, 7000010142, 7000010177, 7000010178, 7000010148, 7000010149, 7000010157, 7000010137, 7000010175, 7000010176, 7000010145, 7000010146, 7000010147, 7000010309, 7000010305, 7000010307, 7000010296, 7000010143, 7000010150, 7000010151, 7000010155. P60034/1000022, 1000023, 1000024, 1000016, 1000017, 1000018, 1000019, 1000020, 1000021
Why was this product recalled?
Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
What actions did Stryker Communications take?
On 7/15/2025, recall notices were mailed or hand delivered to customers who were asked to do the following: Inform individuals within your organization who need to be aware of this action. 2. A Stryker representative will be in contact with your account to schedule a field service visit at your earliest convenience to complete the following: a. Deinstall impacted Oculan light(s) and reinstall with SLX lights to ensure your operating room stays in use. b. Upon product remediation and availability, Stryker will return to reinstall Oculan within your operating room(s). Your local Stryker representative will provide timing updates as they are available. 3. Complete and return the Business Reply Form to [email protected] 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2620-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology