FDA RECALL RECORD · Z-2621-2025

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app) Recall

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

FDA recall Class II

Ongoing

FDA lists this Class II recall for t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2621-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 3, 2025
FDA report date
Oct 1, 2025
Quantity in commerce
19
Distribution
See key details
FDA cause
Software design

What is being recalled?

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

Affected model & UDI checker

Search the exact identifiers published for Z-2621-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedt:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)Check source notice00389152470641
FDA code information

Tandem t:slim Mobile App, Android Software Version 2.9.1 UDI-DI code: 00389152470641 Serial Numbers / Cell Phone OS: 1163923 Google Pixel 8 Pro husky [15 35] 1164430 samsung SM-S901E r0qxxx [15 35] 1167131 samsung SM-A346E a34xdxx [14 34] 1299238 samsung SM-G780G r8qxx [13 33] 1300180 samsung SM-S928B e3qxxx [15 35] 1300180 samsung SM-S928B e3qxxx [14 34] 1329242 Google Pixel 6 Pro raven [15 35] 1381313 samsung SM-G780G r8qxx [13 33] 1381315 Google Pixel 9 tokay [16 36] 1381537 samsung SM-A546E a54xnsxx [14 34] 1381570 Xiaomi 2407FPN8EG rothko_global [15 35] 1391808 samsung SM-S711B r11sxxx [14 34] 1391892 samsung SM-N980F c1sxx [13 33] 1392863 samsung SM-S926B e2sxxx [15 35] 1392953 samsung SM-S916B dm2qxxx [15 35] 1393040 samsung SM-S901E r0qxxx [15 35] 90615975 realme RMX3241 RMX3241 [13 33] 90754587 samsung SM-S916B dm2qxxx [15 35] 901144584 samsung SM-S901E r0qxxx [14 34]

View the original FDA recall notice

Why was this product recalled?

an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Tandem Diabetes Care, Inc. take?

On 07/03/2025, the firm sent an "URGENT FIELD SAFETY NOTICE" it distributor to notify customers that users operating an Android phone with language setting set as Hebrew may experience a blank screen on the mobile application during the initial pairing sequence. The language setting must be set to English for the pairing sequence to work successfully. However, if users set their language back to a right-to-left language after pairing, or do not encounter the initial issue, the mobile application will display discrepancies. This discrepancy may cause a user to make incorrect therapy decisions that may result in under-delivery or over-delivery of insulin. Customers are instructed to: If you are operating the Tandem t:slim mobile app version 2.9.1 on an Android smartphone, and your device language is set to a right-to-left language, please revert your phone language setting to English. You can do this by navigating to the Settings app on your device. If you are unable to or do not want to change your phone language to English, Tandem recommends not to use the Android Tandem t:slim mobile app version 2.9.1 until a future update addresses this issue. Additional Recommended Best Practices: 1. Continue using your Tandem pump as described in the User Guide. 2. Regularly check your blood sugar to ensure you are not having unexpectedly high or low readings. 3. Pay attention to all system alerts and alarms. 4. Always carry back-up supplies. For questions or assistance - Additional information of this Field Safety Notice is available by contacting Padagis at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.