A M S, ALIGNED MEDICAL SOLUTIONS, Dr. Lewin Pack, Pack Number AMS7200D, surgical convenience kit Recall
A M S, ALIGNED MEDICAL SOLUTIONS, Dr. Lewin Pack, Pack Number AMS7200D, surgical convenience kit
FDA recall Class I
Ongoing
FDA lists this Class I recall for A M S, ALIGNED MEDICAL SOLUTIONS, Dr. Lewin Pack, Pack Number AMS7200D, surgical convenience kit. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2622-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 21, 2026
- FDA report date
- Jul 8, 2026
- Quantity in commerce
- 45 units
- Distribution
- See key details
- FDA cause
- Material/Component Contamination
What is being recalled?
A M S, ALIGNED MEDICAL SOLUTIONS, Dr. Lewin Pack, Pack Number AMS7200D, surgical convenience kit
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
UDI/DI none, Lot Codes: 237663, 237901, 237957, 238041, 238084, 238085, 238146, 238197, 238589, 239157, 239216, 239286, 239337, 239386, 239943, 240269, 240633, 240705, 240903, 240904, 241034, 241223, 241224, 241313.
Why was this product recalled?
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
What actions did Windstone Medical Packaging, Inc. take?
AMS ALIGNED MEDICAL SOLUTIONS issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 5/21/2026 via email. The notice explained the issue, potential risk, and requested the following: "AMS requests that you undertake the following activities: 1. Immediately check your inventory for the recalled products listed above and place them under quarantine. 2. Due to the manufacturing process any finished sterilized packs remaining in inventory, or at your facility will require over labeling for all products affected by this recall. Packs will be labeled with a sticker attached to each pack. This sticker will read: RECALL NOTICE CARDINAL HEALTH announced the recall of 5110 Webcol" Large Alcohol Prep Pad contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall. " At the time the kit is opened for use any Cardinal Alcohol Prep Pad should be identified and set aside. " The recalled Cardinal Alcohol Prep Pad should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. 3. Complete the attached Recall Reply Form listing all inventory of affected product by lot number and quantity. 4. If you have further distributed this product, please identify your distribution partners, and notify them at once of the product recall. Your notification to your customers should include a copy of this notification letter. 5. Please complete the Recall Reply Form within 5 business days and return via email to [email protected] even if you have no affected product on hand." For questions or concerns: [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2622-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.