Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2630-2026

CRE Pro Wireguided 15-18mm 240cm Recall

CRE Pro Wireguided 15-18mm 240cm

FDA recall Class II

Ongoing

FDA lists this Class II recall for CRE Pro Wireguided 15-18mm 240cm. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2630-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 21, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
81,543 units
Distribution
See key details
FDA cause
Package design/selection

What is being recalled?

CRE Pro Wireguided 15-18mm 240cm

Affected model & UDI checker

Search the exact identifiers published for Z-2630-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedCRE Pro Wireguided 15-18mm 240cmCheck source notice8714729797586
FDA code information

UPN: M00558700; GTIN: 8714729797586; Lot No. 38043116, 38192116, 38192117, 38192118, 38192119, 38192960, 38192961, 38214896, 38214897, 38222153, 38222496, 38222497, 38222498, 38222499, 38222660, 38222661, 38222662, 38222663, 38222664, 38222665, 38238864, 38239084, 38239085, 38239086, 38239087, 38239088, 38239089, 38241874, 38241875, 38241876, 38242679, 38242820, 38242821, 38242822, 38242824, 38269134, 38269135, 38274558, 38274559, 38282412, 38324631, 38324632, 38324633 ,38324634, 38324636, 38324637, 38324638, 38347325, 38347326, 38347327, 38347328, 38364300, 38381887, 38421906, 38421907, 38421908, 38421909, 38421910, 38421911, 38421912, 38421913, 38421914, 38441933, 38441934, 38441935, 38441936, 38450077, 38450078, 38450079, 38450260, 38450261, 38450262, 38450263, 38462131, 38462132, 38462133, 38468755, 38468756, 38472371, 38472372, 38472373, 38479757, 38479758, 38503030, 38507563, 38507564, 38507565, 38507566, 38507567, 38507568, 38507569, 38568778, 38568779, 38568860, 38585010, 38585011, 38585012, 38585013, 38621957, 38621958, 38642724, 38642725, 38642726, 38642727, 38665032, 38665033, 38665034, 38665035, 38665036, 38665038, 38678315, 38688161, 38726662, 38726663, 38726664, 38734730, 38734731, 38745591, 38748249, 38748250, 38748251, 38748253, 38762224, 38762225, 38762226, 38762227, 38762228, 38769690, 38769691, 38769692, 38800533, 38800534, 38800535, 38800536, 38831782, 38831783, 38831784, 38831785, 38831786, 38831787, 38831788, 38877007, 38877008, 38877009, 38878807, 38878808, 38878809, 38878810, 38878811, 38878812, 38886257, 38886258, 38886259, 38916782, 38916783, 38916784, 38916785, 38933957, 38933958, 38933959, 38935360, 38945672; Exp. November 24, 2028 March 26, 2029.

View the original FDA recall notice

Why was this product recalled?

Potential sterile breach of the pouches in which devices are packaged.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did BOSTON SCIENTIFIC CORPORATION take?

An Urgent Medical Device Removal - Immediate Action Required notification was mailed to consignees on 5/21/26. This notification informed consignees that Boston Scientific was conducting a removal of affected devices due to a potential sterile breach of product packaging (pouches). Consignees are instructed to immediately stop further use or distribution of product and quarantine them. Consignees are to complete and return the provided Reply Verification Tracking Form and return the product to Boston Scientific per the provided instructions. Consignees are to share the recall notification with any healthcare provider that uses the affected devices and with organizations to which product was further distributed. Consignees with any questions are to contact their Boston Scientific representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology