Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2639-2025

ZAP-X Radiosurgery System, Model: 300150 Recall

ZAP-X Radiosurgery System, Model: 300150

FDA recall Class II

Ongoing

FDA lists this Class II recall for ZAP-X Radiosurgery System, Model: 300150. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2639-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 25, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
17
Distribution
See key details
FDA cause
Software change control

What is being recalled?

ZAP-X Radiosurgery System, Model: 300150

Affected model & UDI checker

Search the exact identifiers published for Z-2639-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
300150ZAP-X Radiosurgery System, Model: 300150Check source noticeNot specified
Not specifiedZAP-X Radiosurgery System, Model: 300150Check source notice00860183001504
FDA code information

UDI-DI: 00860183001504, Serial Numbers: ZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181. Software version: TDS V1.10.1

View the original FDA recall notice

Why was this product recalled?

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Zap Surgical Systems take?

On 7/25/2025, correction notices were sent to customers who were asked to do the following: Firm will be revising the software to address this issue. Until we supply you with the revised software, the firm strongly advises that you follow these steps to prevent potential collisions. 1. In the rare event that the system triggers a proximity error message during a long gantry move (greater-than-180-degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, please manually extract the patient. 2. After the patient has been safely extracted, please return the gantry to its home position before resuming any treatment procedures. If you have any further questions related to this situation email [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology