FDA RECALL RECORD · Z-2641-2026

ARTIS icono biplane. Recall

ARTIS icono biplane. Model Number: 11327600.

FDA recall Class II

Ongoing

FDA lists this Class II recall for ARTIS icono biplane.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2641-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 20, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
85 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

ARTIS icono biplane. Model Number: 11327600.

Affected model & UDI checker

Search the exact identifiers published for Z-2641-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11327600ARTIS icono biplane.Check source noticeNot specified
Not specifiedARTIS icono biplane.Check source notice4056869063317
FDA code information

Model Number: 11327600. UDI-DI: 4056869063317. Serial Numbers: 184246, 184241, 184086, 184178, 184108, 184099, 184070, 184127, 184095, 184167, 184184, 184205, 184199, 184221, 184126, 184047, 184093, 184044, 184096, 184214, 184100, 184119, 184073, 184188, 184134, 184150, 184081, 184153, 184085, 184243, 184117, 184045, 184159, 184151, 184177, 184079, 184087, 184169, 184116, 184157, 184163, 184123, 184090, 184118, 184140, 184230, 184224, 184195, 184248, 184225, 184190, 184106, 184046, 184078, 184200, 184142, 184138, 184166, 184152, 184113, 184186, 184196, 184105, 184102, 184147, 184120, 184121, 184124, 184211, 184233, 184216, 184050, 184067, 184112, 184122, 184220, 184089, 184097, 184234, 184103, 184098, 184084, 184183, 184179, 184109

View the original FDA recall notice

Why was this product recalled?

Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Medical Solutions USA, Inc. take?

Siemens notified consignees on about 05/19/2026 via letter. Consignees were instructed to ensure all users of the affected products within each organization and others who may need to be informed will receive the relevant safety information provided within the notice and comply with the recommendations therein, maintain the recall notification with product related records and until the corrective measures have been completed. Siemens will be releasing a software update to correct the issue via an Update Instruction AX018/26/S. All affected systems will be corrected by Siemens Healthineers. The service organization will contact consignees in the near future to arrange a date to perform the corrective action. Additionally, consignees were instructed to confirm and acknowledge receipt of the recall notification digitally.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology