CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling. Recall
CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling.
FDA recall Class II
Ongoing
FDA lists this Class II recall for CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2656-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 21, 2026
- FDA report date
- Jul 8, 2026
- Quantity in commerce
- 3,240
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling.
Affected model & UDI checker
Search the exact identifiers published for Z-2656-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling. | Check source notice | 00801741121609 |
| Not specified | CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling. | Check source notice | 00801741108051 |
FDA code information
CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 Lot Number: REKV1588 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 Lot Number: REKX0938 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 Lot Number: REKY1965
Why was this product recalled?
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
What actions did Bard Access Systems, Inc. take?
On 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the drug product from the BD convenience kits/procedure trays at the point of use, within a controlled and sterile environment to maintain the sterility of the remaining components within the kit. All other kit/tray components are safe to use. 4. If affected Lidocaine ampule was previously used without issue, no additional treatment or clinical follow up is required. 5. If affected Lidocaine ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Additional actions to take: 1. Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2. Post this notice on all inventory storage locations and any location where the product may be used. 3. Label affected kits/trays using the template provided in Appendix 2 to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. a. Apply the labels to the product without covering or obscuring any regulatory or traceability information. b. Apply to the top web of the kit/tray pressing firmly to ensure adhesion, while taking care not to damage or compromise the sterile barrier. For Questions - Contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompt
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2656-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026