4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 4382A SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108501 The Soft Tissue Biopsy devices are indicated for use in the collection of aspirate tissue (e.g. liver, kidney, etc.). Recall

4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 4382A SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108501 The Soft Tissue Biopsy devices are indicated for use in the collection of aspirate tissue (e.g. liver, kidney, etc.).

FDA recall Class II

Ongoing

FDA lists this Class II recall for 4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 4382A SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108501 The Soft Tissue Biopsy devices are indicated for use in the collection of aspirate tissue (e.g. liver, kidney, etc.).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2668-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 21, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
11,990
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 4382A SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108501 The Soft Tissue Biopsy devices are indicated for use in the collection of aspirate tissue (e.g. liver, kidney, etc.).

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 Lot Number: REKY3390 4382A SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108501 Lot Numbers: REKY3393 REJZ0942 REJZ0943

View the original FDA recall notice

Why was this product recalled?

Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

What actions did Bard Access Systems, Inc. take?

On 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the drug product from the BD convenience kits/procedure trays at the point of use, within a controlled and sterile environment to maintain the sterility of the remaining components within the kit. All other kit/tray components are safe to use. 4. If affected Lidocaine ampule was previously used without issue, no additional treatment or clinical follow up is required. 5. If affected Lidocaine ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Additional actions to take: 1. Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2. Post this notice on all inventory storage locations and any location where the product may be used. 3. Label affected kits/trays using the template provided in Appendix 2 to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. a. Apply the labels to the product without covering or obscuring any regulatory or traceability information. b. Apply to the top web of the kit/tray pressing firmly to ensure adhesion, while taking care not to damage or compromise the sterile barrier. For Questions - Contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompt

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology