Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 Recall
Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006
FDA recall Class II
Ongoing
FDA lists this Class II recall for Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2672-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 8, 2025
- FDA report date
- Oct 8, 2025
- Quantity in commerce
- 266 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006
Affected model & UDI checker
Search the exact identifiers published for Z-2672-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 3.921002 | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 | Check source notice | Not specified |
| 3.921030 | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 | Check source notice | Not specified |
| 3.921006 | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 | Check source notice | Not specified |
| Not specified | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 | Check source notice | 07613365001914 |
| Not specified | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 | Check source notice | 07613365003468 |
| Not specified | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 | Check source notice | 07613365003475 |
| Not specified | Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 | Check source notice | 07613365000115 |
FDA code information
REF: 1A.701307: UDI- 07613365003024 REF: 3.921002, UDI: 07613365001914 REF: 3.921030, UDI: 07613365003468 REF: 3.921031: UDI: 07613365003475 REF: 3.921006, UDI: 07613365000115 Affected Serial Numbers: 7010.000160 and lower, 7012.000356 and lower, 7013.000256 and lower, 7014.000152 and lower
Why was this product recalled?
Vital sign monitoring instrument may trigger an error message during blood pressures measurements, Specifically, when the initial inflation pressure is set in the high-pressure range.
What actions did SCHILLER AG take?
On August 8, 2025, Schiller AG issued an Field Safety Notice to affected consignees via E-Mail. On September 6, 2025, Schiller issued a "Urgent Medical Device Recall Expansion their recall to additional products. Schiller asked consignees to take the following actions: 1. Read the Field Safety Notice (FSN) provided by the system manufacturer carefully 2. Wait until your system manufacturer contacts you for device testing or to arrange the remediation of your device(s). 3. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. (As appropriate) 4. Please transfer this notice to other organizations on which this action has an impact. (As appropriate) 5. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2672-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.