Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Recall
Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 342-19-703 Product Description: Enovis knee devices are intended for knee arthroplasty.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2700-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 16, 2026
- FDA report date
- Jul 15, 2026
- Quantity in commerce
- 4
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 342-19-703 Product Description: Enovis knee devices are intended for knee arthroplasty.
Affected model & UDI checker
Search the exact identifiers published for Z-2700-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 342-19-703 | Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE | Check source notice | Not specified |
| Not specified | Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE | Check source notice | 00888912167703 |
FDA code information
Model/Catalog Number: 342-19-703 Lot Code: Lot: 177T1057 GTIN: 00888912167703 Expiration Date: 12/07/2026
Why was this product recalled?
Due to packaging discrepancy with incorrect product size mixed up.
What actions did Encore Medical, LP take?
On 02/16/2026, the firm emailed an "Urgent: Medical Device Recall" letter to customers informing them that Enovis (Encore Medical L.P) became aware that there is a potential packaging discrepancy with the EMPOWR 3D Knee INS. 342-19-703 (Lot 177T1057), EMPOWR 3D Knee INS, 3R, 19MM, VE, was found within the packaging of a 342-14-711 (Lot 145T1067), EMPOWR 3D Knee INS, 11R, 14MM, VE Customers are instructed to: 1. Review stock for the products and lot numbers for the items to be returned. 2. Notice needs to be passed on to all those within their organization or to any organization where the product has been transferred. 3. Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 or through regional email address, to place a replacement request. An RMA number and shipping label will be provided to return the affected product. 4. Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service. For any questions, call/text 682-299-9133 or email [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2700-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology