Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2700-2026

Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Recall

Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 342-19-703 Product Description: Enovis knee devices are intended for knee arthroplasty.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2700-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 16, 2026
FDA report date
Jul 15, 2026
Quantity in commerce
4
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 342-19-703 Product Description: Enovis knee devices are intended for knee arthroplasty.

Affected model & UDI checker

Search the exact identifiers published for Z-2700-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
342-19-703Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VECheck source noticeNot specified
Not specifiedBrand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VECheck source notice00888912167703
FDA code information

Model/Catalog Number: 342-19-703 Lot Code: Lot: 177T1057 GTIN: 00888912167703 Expiration Date: 12/07/2026

View the original FDA recall notice

Why was this product recalled?

Due to packaging discrepancy with incorrect product size mixed up.

What actions did Encore Medical, LP take?

On 02/16/2026, the firm emailed an "Urgent: Medical Device Recall" letter to customers informing them that Enovis (Encore Medical L.P) became aware that there is a potential packaging discrepancy with the EMPOWR 3D Knee INS. 342-19-703 (Lot 177T1057), EMPOWR 3D Knee INS, 3R, 19MM, VE, was found within the packaging of a 342-14-711 (Lot 145T1067), EMPOWR 3D Knee INS, 11R, 14MM, VE Customers are instructed to: 1. Review stock for the products and lot numbers for the items to be returned. 2. Notice needs to be passed on to all those within their organization or to any organization where the product has been transferred. 3. Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 or through regional email address, to place a replacement request. An RMA number and shipping label will be provided to return the affected product. 4. Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service. For any questions, call/text 682-299-9133 or email [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.