TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components Recall
TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components
FDA recall Class II
Ongoing
FDA lists this Class II recall for TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2716-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 8, 2026
- FDA report date
- Jul 15, 2026
- Quantity in commerce
- 360 devices
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components
Affected model & UDI checker
Search the exact identifiers published for Z-2716-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 109003 | TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components | Check source notice | Not specified |
| 109004 | TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components | Check source notice | Not specified |
FDA code information
Model Number: 109003 UDI-DI code: M9461090030 Model Number: 109004 UDI-DI code: M9461090040
Why was this product recalled?
Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did THINK Surgical, Inc. take?
A "MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that THINK Surgical Incorporated is voluntarily notifying you of a potential defect in its Tracking Arrays (Femur and Tibia). THINK Surgical Incorporated has become aware that the tracking arrays may become loose. There have been 11 complaints regarding loose tracking arrays. if a Tracking Array is loose and moves when attached to the Tracking Array Bracket, it could result in incorrect positional information communicated to the system, potentially impacting system accuracy. Customers are instructed to: Prior to each use of any tracking array, the user should confirm that the tracking array is rigidly fixed in place and not loose when attached to the tracking array bracket. If the tracking array is loose, please use a replacement tracking array. -Fill out the accompanying Acknowledgement Form. Email the completed Acknowledgement Form to THINK Surgical, Inc at [email protected]. - Examine their inventory of tracking arrays and inspect them before each use. Please see Appendix 1 attached for instructions to perform the inspection method. - THINK Surgical Incorporated will provide customers with a Returned Material Authorization Form (RMA), a FedEx Envelope and return Fedex AirBill to return any affected devices. Customers will receive replacement unit(s). For Questions - Contact THINK Surgical Incorporated Customer Service at phone 833-878-4465 or [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
| Product / regulatory record | Company | Pathway | Decision date | Source |
|---|---|---|---|---|
| THINK Surgical, Inc. | 510(k) clearance | Mar 3, 2026 | FDA |
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2716-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.