Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2719-2026

Olympus Toomey Evacuator 50cc with Adaptor. Recall

Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Olympus Toomey Evacuator 50cc with Adaptor.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2719-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 5, 2026
FDA report date
Jul 22, 2026
Quantity in commerce
1,156 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.

Affected model & UDI checker

Search the exact identifiers published for Z-2719-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
188AOlympus Toomey Evacuator 50cc with Adaptor.Check source noticeNot specified
Not specifiedOlympus Toomey Evacuator 50cc with Adaptor.Check source notice00821925001077
FDA code information

Model Number: 188A. UDI-DI: 00821925001077. All Lot Numbers

View the original FDA recall notice

Why was this product recalled?

Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on about 06/05/2026 via Courier, 2-day delivery. Consignees were requested to examine inventory for affected units. If affected units are on hand, replace any existing sterilization instructions for these products with those provided with the notification and add these instructions to any existing copies of the IFU. Consignees were also instructed to acknowledge receipt of the customer letter through the web portal. If affected units were further distributed, Olympus instructed consignees to identify those customers and forward the notification to them. An updated notice was sent to consignees on 06/30/2026 with further clarification regarding instructions to be used to sterilize affected devices going forward.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology