da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15 Recall
da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15
FDA recall Class II
Ongoing
FDA lists this Class II recall for da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2722-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 12, 2025
- FDA report date
- Jul 22, 2026
- Quantity in commerce
- 11113
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15
Affected model & UDI checker
Search the exact identifiers published for Z-2722-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15 | Check source notice | 00886874113486 |
FDA code information
UDI-DI: 00886874113486, Lot: 3074LA0500, 1124LA0200, 2124LA0400, 1134LA0700, 1144LA0500, 7154LA0200, 6164LA0800, 1084LA0300, 3164LAA800, 4084LA0500, 4124LA0600, 1274LA0500, 2294LA0200, 1274LA0400, 4284LA0800, 6324LA0200, 1324LA0300, 4154LA0400, 2284LA0100, 4144LA1300, 3324LA0500, 2354LA0500, 7144LA0500, 4294LA0700, 3374LA0400, 5264LA0700, 3224LA0500, 3424LA0200, 1394LA0100
Why was this product recalled?
Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Intuitive Surgical, Inc. take?
On 5/12/2025, correction notices were emailed to customers informing them of the following: 1) Conduct pre-use inspection steps outlined prior to each procedure, see below. 2) If you encounter a drape that does not pass the inspection steps, contact firm customer service. 3) Complete and return the acknowledgement form via email to [email protected] or [email protected] If you need further information or support concerning this Safety Notice, please contact your Clinical Sales Representative or contact firm Customer Service at the numbers listed below: - North and South America: (800) 876-1310, Option 3 or mail: [email protected]. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 or [email protected] - South Korea: 02-3271-3200or [email protected] - Japan: 0120-56-5635 or 03-5575-1362 or [email protected] " Taiwan: +0800-86-8181 Pre-Use Inspection Steps: 1. Fully drape the da Vinci SP system according to user manual. 2. Press and hold the Instrument Arm Clutch button. Top Instrument arm clutch, not recommended. 3. While holding the Instrument Arm Clutch button, manually rotate the instrument drive cluster in one direction, 360-degrees. Then, rotate the instrument drive at least 360-degrees in the other direction. Stop immediately and press "deploy for draping" again to rehome the instrument drive cluster to replace the drape if you observe any of the following behaviors: a. Inability to rotate the instrument drive 360-degrees in both directions b. VSC message that drape is not fully attached (VSC message: "Check drape is secure to all 4 instrument arm clips" or "Insecure or missing drape. The drape may not be properly secured and could fall. Ensure the drape is properly connected to all 4 mounts to continue"). c. Drape outer ring falls off silver section of system. d. Any of the 4 dark gray marks disengage from the dark gray connectors during the rotation.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2722-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology