FDA RECALL RECORD · Z-2726-2026

Optima XR200amx, Mobile Digital-Ready Radiographic System Recall

Optima XR200amx, Mobile Digital-Ready Radiographic System

FDA recall Class II

Ongoing

FDA lists this Class II recall for Optima XR200amx, Mobile Digital-Ready Radiographic System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2726-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 5, 2026
FDA report date
Jul 22, 2026
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Optima XR200amx, Mobile Digital-Ready Radiographic System

Affected model & UDI checker

Search the exact identifiers published for Z-2726-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedOptima XR200amx, Mobile Digital-Ready Radiographic SystemCheck source notice00840682115278
FDA code information

UDI-DI: 00840682115278; System IDs: 419383DP3, 773792XR200, 856355OPT, FEB520100

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems, LLC take?

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 6/5/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use the system by following the instructions below: Perform a check of your device prior to each patient exam to look for the following behaviors: "Column vertical movement requires increased effort. "Column has change in resistance (i.e. movement is not smooth) during vertical movement. "Column makes abnormal noises during vertical movement. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 10967 Digital Response Form or print, fill out manually, scan, and email to [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.