Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface Recall
Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface Model/Catalog Number: RII1001 (Small), RII1002 (Medium), RII1003 (Large) Software Version: N/A Product Description: Nasal pillows interface intended for use with positive airway pressure therapy devices. The device delivers airflow from a PAP device to the patient's nose and is secured using adjustable headgear. Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2740-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 15, 2023
- FDA report date
- Jul 22, 2026
- Quantity in commerce
- 375
- Distribution
- See key details
- FDA cause
- Labeling Change Control
What is being recalled?
Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface Model/Catalog Number: RII1001 (Small), RII1002 (Medium), RII1003 (Large) Software Version: N/A Product Description: Nasal pillows interface intended for use with positive airway pressure therapy devices. The device delivers airflow from a PAP device to the patient's nose and is secured using adjustable headgear. Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-2740-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| RII1001 | Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface | Check source notice | Not specified |
| RII1002 | Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface | Check source notice | Not specified |
| RII1003 | Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface | Check source notice | Not specified |
FDA code information
Lot Code: Affected Catalog Numbers: RII1001, RII1002, RII1003 Affected Lots RII1001: 220121 220121 220191 RII1002: 220121 220121 220191 RII1003: 220121 220121 220191
Why was this product recalled?
Packaging and labeling discrepancy involving product size identification and associated Unique Device Identifier (UDI/GTIN).
What actions did 3B Medical, Inc. take?
This is a retroactively reported action. On June 22, 2023, the firm notified affected customers via letter titled "Urgent Medical Device Recall." Customers were informed of the incorrect labeling. Customers were asked to examine their inventory to determine if they had any of the affected products on-hand. Customers were to discontinue distribution of identified products and to destroy any affected products on hand. Firm offered to issue credit memo. If product was further distributed, customers were asked to notify patients/downstream accounts and have the product returned for disposal.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2740-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology