The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274. Recall
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
FDA recall Class II
Ongoing
FDA lists this Class II recall for The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2741-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 2, 2026
- FDA report date
- Jul 22, 2026
- Quantity in commerce
- 205 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
Affected model & UDI checker
Search the exact identifiers published for Z-2741-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| VH-4000 | The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274. | Check source notice | Not specified |
| Not specified | The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274. | Check source notice | 00607567700406 |
FDA code information
UDI 00607567700406, Serial/Lot Number 3000547797
Why was this product recalled?
Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
What actions did Maquet Cardiovascular, LLC take?
Getinge issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/18/2026 via FedEx. The notice explained the issue, risk to health, and requested the following actions be taken: " Immediately identify and quarantine all affected Vasoview Hemopro 2 devices from lot 3000547797. " Do not use affected devices. " If an affected device was used, no further intervention is required. Standard post-operative wound care per facility protocol should be followed. " If any affected devices were transferred within your organization or to another organization, ensure this notice is shared with the appropriate individuals. " Contact Getinge Customer Service 1-888-880-2874 or your Getinge representative to coordinate return of the affected units and replacement product. " Complete and return the enclosed response form to Getinge by e-mailing a copy to [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2741-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology