FDA RECALL RECORD · Z-2746-2026

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204 Recall

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204

FDA recall Class II

Ongoing

FDA lists this Class II recall for ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2746-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 12, 2026
FDA report date
Jul 22, 2026
Quantity in commerce
310 units
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204

Affected model & UDI checker

Search the exact identifiers published for Z-2746-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
0025204ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204Check source noticeNot specified
Not specifiedST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204Check source notice14560189210032
FDA code information

UDI-DI: 14560189210032; Lot Numbers: FX12404, FY12407, FY12406, FZ12408

View the original FDA recall notice

Why was this product recalled?

The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.

What actions did Tosoh Bioscience Inc take?

Tosoh Bioscience Inc. sent an URGENT: MEDICAL DEVICE RECALL letter dated 6/9/26 to customers. Customers are to SEGREGRATE and quarantine all affected lots in inventory. Affected product lot should not be used. CONTACT Tosoh Customer Service department at 1-866-527-3587 to arrange return of the affected product. For questions, contact Tosoh Technical Support [email protected] or 1-800-248-6764. For replenishment of returned Testosterone test cups, contact Customer Service department at 1-866-527-3587. Replenishments will be with ST AIA-PACK Testosterone Assay Test Cups from new lots. If you have obtained unexpected test results or received any complaints of illness or adverse events associated with the use of the ST AIA-PACK Testosterone Assay Test Cups and associated Testosterone reagents, contact Tosoh Technical Support at 1-800-248-6764 or [email protected] Monday to Thursday 8:30AM-8:30PM EST Friday 8:30AM-7:30PM EST An URGENT: MEDICAL DEVICE RECALL: SECOND NOTIFICATION: dated 7/17/26 was sent to customers. The letter has additional language regarding the risks to health associated with the use of the affected lots. Instructions to customers are the same as in the first letter.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology