Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2762-2026

SeaSpine Polyaxial Head REF Number: 41-3010 Recall

SeaSpine Polyaxial Head REF Number: 41-3010

FDA recall Class II

Ongoing

FDA lists this Class II recall for SeaSpine Polyaxial Head REF Number: 41-3010. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2762-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 28, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
796 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

SeaSpine Polyaxial Head REF Number: 41-3010

Affected model & UDI checker

Search the exact identifiers published for Z-2762-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSeaSpine Polyaxial Head REF Number: 41-3010Check source notice10889981155117
FDA code information

UDI: 10889981155117; Affected Lots: AW199682G, AW199684G.

View the original FDA recall notice

Why was this product recalled?

Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.

What actions did SeaSpine Orthopedics Corporation take?

On June 16, 2026, Orthofix issued a updated Urgent Medical Device Recall notification to provide additional information concerning the recall. Initial notification was issued on May 28, 206. Orthofix ask consignees to take the following actions: 1. Discontinue use of affected lots, 2.Immediately identify and quarantine any affected product. 3. If product is identified, Orthofix will facilitate the return. 4. Complete and return the attached Acknowledgement and Receipt Form. 5. This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology