Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2763-2026

Halyard EYE PACK. Recall

Halyard EYE PACK. Model/Catalog Number: LVEY27 Convenience kit.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Halyard EYE PACK.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2763-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 7, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
260 kits
Distribution
See key details
FDA cause
Process control

What is being recalled?

Halyard EYE PACK. Model/Catalog Number: LVEY27 Convenience kit.

Affected model & UDI checker

Search the exact identifiers published for Z-2763-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
LVEY27Halyard EYE PACK.Check source noticeNot specified
Not specifiedHalyard EYE PACK.Check source notice10809160499528
FDA code information

Model/Catalog Number: LVEY27. UDI-DI: 10809160499528. Lot Numbers (Expiration Date): 1680076 (12/31/2026), 1680590 (6/30/2027).

View the original FDA recall notice

Why was this product recalled?

Incorrect concentration of Betadine solution contained within certain AVID convenience kits.

What actions did AVID Medical, Inc. take?

AVID Medical notified the single consignee on about 05/07/2026 via email. The consignee was instructed to identify, segregate, and quarantine all impacted products, add the provided warning labels to affected AVID kits to ensure all users are aware of the issue, and discard all impacted components upon opening the Eye Pack. Customers and end users were instructed that if they received affected products from a distributor, to please contact that distributor of sales representatives with questions. Distributors were instructed that if any of the affected kits identified were further distributed, to ensure all end-users are appropriately notified and maintain records of effectiveness. Recipients were requested to complete and return the provided Recall Response Form by 05/22/2026.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology