Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets Recall
Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets
FDA recall Class I
Ongoing
FDA lists this Class I recall for Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2766-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Jun 30, 2026
- FDA report date
- Aug 19, 2026
- Quantity in commerce
- 24528 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets
Affected model & UDI checker
Search the exact identifiers published for Z-2766-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets | Check source notice | 00085412014319 |
FDA code information
UDI/DI 00085412014319, Lot Numbers: DR26C27062, exp. 30-MAR-2028; DR26D14031, exp. 15-Apr-2028; DR26E13055, exp. 14-May-2028
Why was this product recalled?
Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.
What actions did Baxter Healthcare Corporation take?
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/30/2026 via USPS, first class mail. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Immediately locate, isolate, and cease all use of the affected lot numbers of the product. The product code and lot number can be found on the individual product package and shipping carton. 2. Contact Baxter Healthcare Center for Service to arrange for return and replacement product at 888-229-0001 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you received this communication directly from Baxter, please acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please contact them to arrange for return of the affected product. Please note that responding to Baxter is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to the director of nursing, director of pharmacy, director of purchasing, facility risk manager, and any other departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of the affected product that you distributed to customers and check the associated box on
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2766-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.