FDA RECALL RECORD · Z-2767-2026

3mensio Workstation (Vascular Fenestrated) software Recall

3mensio Workstation (Vascular Fenestrated) software

FDA recall Class II

Ongoing

FDA lists this Class II recall for 3mensio Workstation (Vascular Fenestrated) software. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2767-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 7, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
165
Distribution
See key details
FDA cause
Software change control

What is being recalled?

3mensio Workstation (Vascular Fenestrated) software

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5

View the original FDA recall notice

Why was this product recalled?

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

What actions did PIE Medical Imaging B.V. take?

On 5/7/2026, correction notices were emailed to customers who were asked to do the following: 1) Do not perform measurements on Philips 3D echo datasets containing the doppler information using 3mensio Workstation 10.7, 10.7 SP1 SP5. 2) Ensure that this notice is distributed to all relevant personnel within your organization and to any organization where the affected product may have been transferred to or is in use. 3) PMI recommends follow-up to verify if results of ordered devices are correctly analyzed/calculated. 4) Complete and return the enclosed Customer Acknowledgment Form and return via email to [email protected] This issue has been resolved in 3mensio Workstation 10.7 SP6. PMI will support users in upgrading to this version.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology