Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2772-2026

Halyard STANDARD LINE INSERTION PACK. Recall

Halyard STANDARD LINE INSERTION PACK. Model/Catalog Number: PVHM234-01. Convenience kit.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Halyard STANDARD LINE INSERTION PACK.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2772-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 19, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
112 kits
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Halyard STANDARD LINE INSERTION PACK. Model/Catalog Number: PVHM234-01. Convenience kit.

Affected model & UDI checker

Search the exact identifiers published for Z-2772-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PVHM234-01Halyard STANDARD LINE INSERTION PACK.Check source noticeNot specified
Not specifiedHalyard STANDARD LINE INSERTION PACK.Check source notice10809160416761
FDA code information

Model/Catalog Number: PVHM234-01. UDI-DI: 10809160416761. Lot Numbers (Expiration date): 1597314 (8/31/2026), 1599022 (5/31/2026).

View the original FDA recall notice

Why was this product recalled?

Defects in the silicone seal, which may result in "doming" of the silicone seal due to improper sealing, potentially leading to fluid occlusion or leakage.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AVID Medical, Inc. take?

AVID Medical notified consignees on about 05/19/2026 via email. Consignees were instructed to identify, segregate, and quarantine all impacted kits, and add the provided warning labels to affected AVID kits to ensure all users are aware of discarding the impacted component prior to use. Distributors were instructed to ensure all end-users are properly notified if affected kits have been further distributed. Customers/End-users were instructed that if they have received product from a distributor, to contact the distributor or your sales representative with any questions. Recipients were requested to complete and return the provided Recall Response Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.