MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android) Recall

MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

FDA recall Class II

Ongoing

FDA lists this Class II recall for MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2778-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 16, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
3,465
Distribution
See key details
FDA cause
Software design

What is being recalled?

MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

Affected model & UDI checker

Search the exact identifiers published for Z-2778-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MMT-8600MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)Check source noticeNot specified
FDA code information

Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0

View the original FDA recall notice

Why was this product recalled?

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medtronic MiniMed, Inc. take?

On 6/16/2026 In-App Push Notifications were sent to customers entitled, Important Field Safety Corrective Action required on your MiniMed Go app. informing them a software upgrade is available to ensure important nighttime notifications are received. Correction notices were also emailed to customers informing them to update to version 1.2.0 or higher. For iOS Users: To confirm the latest app version on your iOS device, follow the steps below: 1. Open the App Store. 2. Tap your profile icon at the top of the screen. 3. Scroll to see pending updates. 4. Tap Update next to the MiniMed Go app icon. For Android Users: To confirm the latest app version on your Android device, follow the steps below: 1. Open the Google Play Store. 2. Tap Menu, then select Manage apps & device. 3. Apps with an update available are labeled "Update available." You can also search for the MiniMed Go app. 4. Tap Update next to the MiniMed Go app icon. Acknowledge that you have read and understood the notification by clicking the acknowledgement button or visiting https://www.minimed.com/en-us/safety-notices/notice31-letter If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

Product / regulatory recordCompanyPathwayDecision dateSource
Medtronic Minimed510(k) clearanceJan 8, 2026FDA

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology