AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Neuro Pack, Pack Number AMS12195C; 2) Neuro Pack, Pack Number AMS12195D; 3) Neuro Pack, Pack Number AMS12307; 4) Neuro Spine Pack, Pack Number AMS16313; 5) Lami Kukkar Pack, Pack Number AMS17958; 6) Neuro Pack, Pack Number AMS6452B; 7) Neuro Lami Pack, Pack Number AMS9427C. Recall
AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Neuro Pack, Pack Number AMS12195C; 2) Neuro Pack, Pack Number AMS12195D; 3) Neuro Pack, Pack Number AMS12307; 4) Neuro Spine Pack, Pack Number AMS16313; 5) Lami Kukkar Pack, Pack Number AMS17958; 6) Neuro Pack, Pack Number AMS6452B; 7) Neuro Lami Pack, Pack Number AMS9427C.
FDA recall Class II
Ongoing
FDA lists this Class II recall for AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Neuro Pack, Pack Number AMS12195C; 2) Neuro Pack, Pack Number AMS12195D; 3) Neuro Pack, Pack Number AMS12307; 4) Neuro Spine Pack, Pack Number AMS16313; 5) Lami Kukkar Pack, Pack Number AMS17958; 6) Neuro Pack, Pack Number AMS6452B; 7) Neuro Lami Pack, Pack Number AMS9427C.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2801-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 8, 2026
- FDA report date
- Aug 5, 2026
- Quantity in commerce
- 927 units
- Distribution
- See key details
- FDA cause
- Material/Component Contamination
What is being recalled?
AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Neuro Pack, Pack Number AMS12195C; 2) Neuro Pack, Pack Number AMS12195D; 3) Neuro Pack, Pack Number AMS12307; 4) Neuro Spine Pack, Pack Number AMS16313; 5) Lami Kukkar Pack, Pack Number AMS17958; 6) Neuro Pack, Pack Number AMS6452B; 7) Neuro Lami Pack, Pack Number AMS9427C.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
1) Pack Number AMS12195C, UDI B098AMS12195C0, Lot Numbers: 217757, 222686, 224418, 225587, 227439, 228598, 230569, 234138, 235331, 236869, 238356. 2) Pack Number AMS12195D, UDI B098AMS12195D0, Lot Numbers: 240974. 3) Pack Number AMS12307, UDI B098AMS12307, Lot Numbers: 220948, 234027, 234756, 235909, 238000, 239036, 241680. 4) Pack Number AMS16313, UDI B098AMS163130, Lot Numbers: 241787. 5) Pack Number AMS17958, UDI B098AMS179580, Lot Numbers: 235956, 237474. 6) Pack Number AMS6452B, UDI B098AMS6452B0, Lot Numbers: 212195, 217806, 220779, 222520, 223506, 224581, 225839, 226652, 227717, 233855, 234503, 236416, 240751. 7) Pack Number AMS9427C, UDI B098AMS9427C0, Lot Numbers: 216097, 216766, 217573, 219221, 222013, 222070, 223714, 225092, 226568, 227813, 230393, 231091, 233313, 236502, 238280, 242405.
Why was this product recalled?
The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Windstone Medical Packaging, Inc. take?
Aligned Medical Solutions (AMS) issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 6/8/2026 via email. The notice explained the issue, potential risk, and requested the following actions: identify and quarantine affected product, affix the sticker which reads: "RECALL NOTICE Medicom announced the recall of Ritmed Neurological Sponges contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall." " At the time the kit is opened for use any Medicom-Ritmed Neurological Sponges should be identified and set aside. " The recalled Medicom-Ritmed Neurological Sponges should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Complete the recall reply form and return it via email to: [email protected] even if you have no affected product on hand. Those further distributing the product were directed to notify their distribution partners. For questions contact [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2801-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.