Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2805-2026

M5 Corpus Power Wheelchair REF,120032 Recall

M5 Corpus Power Wheelchair REF,120032

FDA recall Class II

Ongoing

FDA lists this Class II recall for M5 Corpus Power Wheelchair REF,120032. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2805-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 19, 2025
FDA report date
Aug 5, 2026
Quantity in commerce
715 wheelchairs
Distribution
See key details
FDA cause
Process design

What is being recalled?

M5 Corpus Power Wheelchair REF,120032

Affected model & UDI checker

Search the exact identifiers published for Z-2805-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedM5 Corpus Power Wheelchair REF,120032Check source notice17330818334562
FDA code information

Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562

View the original FDA recall notice

Why was this product recalled?

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

What actions did Permobil Inc take?

On June 10, 2025 Permobil issued a technical bulletin (recall notice) via Email to affected consignees. Permobil informed consignees of the following: 1. Cellular connection is required to receive and install the firmware update. 2. It is important to not use the wheelchair when the firmware update is in process (as indicated on the display of the input device). 3. Specific Remote Update Instructions are available if needed in addition to the on-screen information. 4. A confirmation of the completed update will be shown on the joystick display when the update successfully completes. 5. If you have any questions or concerns, please contact the Permobil Customer Experience team at (800) 736-0925. 1.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology