FDA lists this Class II recall for OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2820-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 2, 2026
FDA report date
Aug 5, 2026
Quantity in commerce
98
Distribution
See key details
FDA cause
Process control

What is being recalled?

OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095

Affected model & UDI checker

Search the exact identifiers published for Z-2820-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
56800095OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095Check source noticeNot specified
02-8000-095OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095Check source noticeNot specified
FDA code information

REF/UDI-DI/Lots: 56800095/10889981149352/181559, 181550 02-8000-095/10889981149314/181559, 181550

View the original FDA recall notice

Why was this product recalled?

Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm configuration) instead of approx. 7.5 cc DBM (fill for 45 ¿ 17.5 mm configuration), which may lead to delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth.

What actions did IsoTis OrthoBiologics, Inc. take?

On 7/2/2026, recall notices were emailed to customers informing them of the following: Orthofix does not recommend removal or revision of implanted devices containing product from the affected lots. Patients should continue to be managed according to the treating surgeon's standard clinical practice and routine follow-up schedule. Surgeons should monitor for signs or symptoms consistent with delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth and manage patients as clinically indicated. If the procedure was performed in accordance with the surgical technique and the intended intraoperative objectives were achieved, no additional patient management beyond routine postoperative care and clinical follow-up described in the Instructions for Use is recommended. The standard postoperative precautions described in the Instructions for Use, including activity restrictions and weight-bearing recommendations, should continue to be followed at the discretion of the treating surgeon. 1) Identify, and quarantine unused product and work with firm to facilitate product return. 2) Complete and return the attached Acknowledgement and Receipt Form 3) This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. Any product complaints should be reported to Orthofix at [email protected]. For questions regarding this recall, including product return, replacement, or credit, please contact Audrey Mudderman, Senior Post Market Surveillance Specialist, at [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology