Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Recall
Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Stryker Hoffmann II Connecting Rod Micro 3 x 40mm.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2828-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 1, 2026
- FDA report date
- Aug 5, 2026
- Quantity in commerce
- 6,244 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.
Affected model & UDI checker
Search the exact identifiers published for Z-2828-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 50796040. | Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. | Check source notice | 04546540389909 |
| 50796040 | Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. | Check source notice | Not specified |
FDA code information
Catalog Number: 50796040. GTIN: 04546540389909. All lot numbers.
Why was this product recalled?
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did STRYKER GmbH take?
Firm began notifying consignees on July 1, 2026. Customers were instructed to immediately check their internal inventory to locate affected product. Customers were informed that until further notice, the subject devices are limited to single use only. Following first use, discard the devices in accordance with the facility's established procedures. Users should conduct visual inspection of product prior to each use. If cracking is found to be present, the customer should quarantine non-confirming parts and report the issue to Stryker's Trauma & Extremities division. Rods with cracks should be returned as part of the complaint process.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2828-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology