REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections Recall
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
FDA recall Class II
Ongoing
FDA lists this Class II recall for REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2834-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 29, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 20 systems
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
Affected model & UDI checker
Search the exact identifiers published for Z-2834-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections | Check source notice | 15099590802073 |
FDA code information
Software Version 2.5 UDI-DI code: 15099590802073
Why was this product recalled?
Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
What actions did Beckman Coulter Biomedical GmbH take?
On 04/29/2026, an "UREGENT MEDICAL DEVICE RECALL" Letter was sent via email and postal informing customers that, Beckman Coulter has identified a defect in Remisol Advance Software version 2.5 when the system is configured with the Generic-Astm Driver. Combo test results are not correctly reported by the software. The issue affects the following assays run on DxI9000 analyzers: "Access HIV Ag/Ab combo (PN C39450) "Access HBsAg Confirmatory (US PN D22915; OUS PN C39425) Customers are instructed to: If you are using Remisol Advance version 2.5: "Immediately review your system configuration (instrument driver for the DxI 9000) and verify that the driver is set to Legacy-Astm (please refer to Annex 1 of this letter for information on how to verify the driver settings); a-If driver is set to Legacy-Astm, no further action is needed; b-If the driver is not set as Legacy-Astm: -Stop immediately the auto-validation of combo tests in Remisol and contact your Beckman Coulter representative or technical support for further instructions and assistance. -Review previously released HIV Ag/Ab combo and HBsAg Confirmatory results, and verify that all sub-results were available for the final result. If they are not available, review the corresponding records on the DxI 9000 Test Results page to identify potentially impacted results. -If the relevant sample data are not available on the DxI 9000 Test Results page, the laboratory should consult with the ordering clinician to determine whether specimen recollection and repeat testing are clinically appropriate. -Follow laboratory procedures to amend reports as needed, if an incorrect result was issued. "Beckman Coulter recommends posting this letter on or near the affected analyzers. Beckman Coulter is targeting Q3 2026 for release software update 2.6 to correct this issue. For questions - contact Customer Support Center from website http://www.beckmancoulter.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2834-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.