Brand Name: WASPLab Product Name: WASPLab Recall
Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: WASPLab Product Name: WASPLab. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2854-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 16, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 1
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.
Affected model & UDI checker
Search the exact identifiers published for Z-2854-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| W087 | Brand Name: WASPLab Product Name: WASPLab | Check source notice | Not specified |
FDA code information
Lot Code: WebApp version 2.33.1, 2.33.3 and 2.34.2
Why was this product recalled?
Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Copan WASP take?
On 06/16/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was provided to customers informing them that, A software bug has been identified in the WebApp software which, under specific conditions, may prevent the result suggested by PhenoMATRIX from being overwritten by the result selected by the user. As a consequence, the result transmitted by WASPLab to the LIS may be the result suggested by PhenoMATRIX, rather than the result selected by the user. Customers were informed and instructed: In date 2026/06/15 we released a WebApp software version including the bugfix of this issue. As a consequence, you are required to schedule with your distributor the software update to correct the issue and, in the meanwhile, to check the results sent to LIS in case you changed the result suggested by PhenoMATRIX. Complete the attached Customer Response Form and return to the Copan WASP Srl contact noted on the form so that Copan WASP Srl may acknowledge your receipt of this notification. For questions or further assistance email [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2854-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology