Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2856-2026

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System Recall

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

FDA recall Class II

Ongoing

FDA lists this Class II recall for munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2856-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 19, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
19
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

Affected model & UDI checker

Search the exact identifiers published for Z-2856-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedmunevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control SystemCheck source notice04260691380010
FDA code information

Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699

View the original FDA recall notice

Why was this product recalled?

A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.

What actions did MUNEVO GMBH take?

On June 19, 2026, the firm issued an "Urgent Medical Device Recall" letter to consignees, which detailed the hazard, mandatory boot-up safety protocols, and manual override steps. The firm has developed software patch Version 12.1.7 (v1), and 12.2.6 (R-net) (released patch on June 17 2026) which corrects the profile retention database loop, ensuring the device consistently locks onto the last saved user profile during boot. This software patch was released to all users on June 17, 2026. Customers should update the software as soon as possible. Until the update is installed, do not activate neutral calibration in the admin menu (only applicable to those with access to the admin menu.)

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.