Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Recall
Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2857-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 11, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 77
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.
Affected model & UDI checker
Search the exact identifiers published for Z-2857-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MC2 | Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System | Check source notice | Not specified |
| MC2.1 | Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System | Check source notice | Not specified |
| Not specified | Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System | Check source notice | 00860008892782 |
| Not specified | Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System | Check source notice | 00850079606043 |
FDA code information
Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547
Why was this product recalled?
Due to a defective circuit board that overheats
What actions did Oxos Medical Inc take?
On 06/12/2026, an "URGENT: Medical Device Recall - Limited to Specific MC2 Units" letter was emailed to customers informing them that, due to faulty circuit boards limited units may experience localized overheating. As a result the firm released a software update (version 5.7.0) which will place a unit in a safe state if boards fail. Customers are instructed to: 1. Update units to software version 5.7.0 via over-air updates. Note, it is important to apply the software update to continue using the MC2 units 2. In addition to the software update, OXOS will replace the defective components with the cassette as part of a scheduled procedure. The MC2 must be returned to OXOS headquarters for replacement of affected components. Servicing will coordinate with customers to minimized any inconvenience. 3. Customer need to acknowledge receipt of the recall communication and complete the software update by calling OXOS Support Team at 866-749-4229 to schedule a time to perform the updated by emailing support#oxos.com, or accepting the update on their devices. For questions or assistance - contact Support Team at 866-749-4229 or email [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2857-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology