Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Recall
Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2858-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 21, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 3740 kits (180 cases)
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791
Affected model & UDI checker
Search the exact identifiers published for Z-2858-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 78791 | Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence | Check source notice | Not specified |
| Not specified | Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence | Check source notice | 20809160242558 |
FDA code information
Lot Code: Lot No.: 327075 328690 319980 320965 318461 322375 325653 330775 331356 336680 UDI: 20809160242558
Why was this product recalled?
Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Medical Action Industries, Inc. 306 take?
On May 21, 2026 Medical Action Industries issued a "Urgent: Medical Device" Recall notification via Email. Medical Action Industries asked consignees to take the following actions: 1. Identify, Segregate, and Quarantine all impacted kits listed in Table 1. 2. Add warning labels to affected AVID kits to ensure all users are aware of discarding the impacted component prior to use. 3. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 4. Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6. Complete the attached Recall Response Form and return by June 5, 2026.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2858-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology