FDA RECALL RECORD · Z-2861-2026

Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Recall

Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2861-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 25, 2025
FDA report date
Aug 12, 2026
Quantity in commerce
2
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.

Affected model & UDI checker

Search the exact identifiers published for Z-2861-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SG-0002Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR InstrumentCheck source noticeNot specified
Not specifiedBrand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR InstrumentCheck source notice00850063063005
FDA code information

Lot Code: SN: C01250010; UDI: (01)00850063063005(11)250218(21)C01250010; SW: 6.0.2 SN: C02250006; UDI:(01)00850063063005(11)250225(21)C02250006; SW: 6.0.2

View the original FDA recall notice

Why was this product recalled?

Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Nuclein, LLC take?

On 06/25/2025, a "MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that, Nuclein identified an incorrect software file on a small number of DASH instruments that creates a possibility for the affected instruments to make incorrect calls that indicate coinfection in the patient tested. On 07/16/2026, the firm emailed an updated "URGENT: MEDICAL DEVICE RECALL" Letter. -Customer were informed regarding updates to the reason for the recall and potential health risks. -Customer were instructed to: All affected DASH Instrument identified in Section 6 (below) have already been replaced. Please complete the Acknowledgement and Receipt Form below and either email the completed form to [email protected] or mail to the address provided at the bottom of the Acknowledgement and Receipt Form. For Questions - contact Nuclein at Phone: (888) 992-DASH on Monday through Friday, 8:30 AM to 5:30 PM, CST. Email: [email protected] Website: https://www.Nuclein.com/contact-us/

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.