BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK Recall
BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK
FDA recall Class I
Ongoing
FDA lists this Class I recall for BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2872-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Jul 30, 2026
- FDA report date
- Aug 26, 2026
- Quantity in commerce
- 80,200
- Distribution
- See key details
- FDA cause
- Component change control
What is being recalled?
BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK
Affected model & UDI checker
Search the exact identifiers published for Z-2872-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| D015151NK | BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK | Check source notice | Not specified |
| D015251NK | BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK | Check source notice | Not specified |
| D015351NK | BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK | Check source notice | Not specified |
| D015451NK | BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK | Check source notice | Not specified |
| D015551NK | BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK | Check source notice | Not specified |
FDA code information
REF/UDI-DI/Lot(Expiration): D015151NK/00801741163586/BSLC9275(2026-08-03), BSLC9563(2026-08-13), BSLC9567(2026-08-25), BSLC10103(2026-09-27), BSLC10516(2026-11-01), BSLC10526(2026-11-16), BSLC10872(2026-12-09), BSLC11216(2028-07-02), BSLC11456(2028-07-23); D015251NK/00801741163678/BSLC9467(2026-08-10), BSLC9575(2026-09-20), BSLC10105(2026-10-04), BSLC10522(2026-11-07), BSLC10868(2026-12-02), BSLC11212(2028-06-18), BSLC11454(2028-07-17), BSLC11463(2028-08-07), BSLC11884(2028-09-15), BSLC11890(2028-09-18), BSLC12082(2028-10-08), BSLC12115(2028-10-20), BSLC12123(2028-10-22); D015351NK/00801741163593/BSLC11458(2028-07-24); 00801741163678/BSLC9277(2026-08-04), BSLC9485(2026-08-12), BSLC9569(2026-08-26), BSLC10099(2026-09-07), BSLC10514(2026-10-16), BSLC10518(2026-10-21), BSLC10524(2026-11-08), BSLC10870(2026-12-03), BSLC10874(2028-06-12), BSLC11218(2028-07-07); D015451NK/00801741163609/BSLC9279(2026-08-06), BSLC9565(2026-08-17), BSLC9571(2026-09-07), BSLC10101(2026-09-17), BSLC10456(2026-10-14), BSLC10520(2026-10-22), BSLC10866(2026-11-21), BSLC11214(2028-06-23), BSLC11450(2028-07-09), BSLC11460(2028-07-31), BSLC11873(2028-08-13), BSLC11888(2028-09-25), BSLC12007(2028-10-02), BSLC12105(2028-10-15); D015551NK/00801741163616/BSLC9465(2026-08-07), BSLC9573(2026-09-26), BSLC10107(2026-10-02), BSLC10528(2026-11-20), BSLC11452(2028-07-10)
Why was this product recalled?
After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Bard Access Systems, Inc. take?
On 7/30/2026, recall notices were mailed/emailed: If rotation and locking occurs, the clinician may need to obtain a new IO needle or other form of vascular access device, such as a Peripheral IV catheter. Alternate vascular access approaches, including additional peripheral IV attempts, or central venous access may continue to be pursued based on clinical judgment/protocols. Affected product should not be used to establish IO access. For patients in whom potentially affected intraosseous needles are currently in use/or have been used and are/were functioning properly, no additional steps need to be taken. 1. Destroy all affected needle set kits, or quarantined until disposal. 2. Provide a copy of this notification to any customers and/or facilities you may have distributed devices to. 3. Complete and return the attached Customer Response Form via email to [email protected] Distributors were also asked to do the following: - A customer letter template is provided to create your own letter to be sent to all customers. - Provide product replacement for all product destroyed by your customer. Questions contact firm at 1-844-823-5433 or [email protected] Firm has acquired new molds with improved dimensional controls. Instructions For Use will be updated with the following to optimize needle / obturator(stylet) / drill handling: Do not pull back on the drill and needle, release the trigger prematurely, or rock/bend the needle during insertion. These actions may result in the needle and drive adapter hubs to rotate and lock together resulting in a bent needle, difficult stylet removal, need for a new device, and/or delay in therapy. Warning: Do not rotate or try to unscrew the drive adapter hub while removing the stylet. These actions may cause the needle and drive adapter hubs to rotate and lock together, resulting in difficult stylet removal, need for a new device, and/or delay in therapy.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2872-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA device-recall notice
First stored snapshot of this source.
- FDA enforcement data
First stored snapshot of this source.