Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2876-2026

GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph Recall

GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2876-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 29, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
163 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

Affected model & UDI checker

Search the exact identifiers published for Z-2876-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGE Healthcare MAC VU360, REF# part number 2030361-023; ElectrocardiographCheck source notice00198953107263
FDA code information

GTIN 00198953107263, Serial Numbers: VU221150078SA, VU222400121SA, VU223230048SA, VU222520132SA, VU219400033SA, VU219400034SA, VU219400045SA, VU219400055SA, VU220270027SA, VU220510048SA, VU220510078SA, VU221100161SA, VU221100181SA, VU221150116SA, VU222120212SA, VU222120220SA, VU222120239SA, VU222120243SA, VU222120249SA, VU222120250SA, VU222120253SA, VU222120258SA, VU222120260SA, VU222120261SA, VU222120266SA, VU222120269SA, VU222140004SA, VU222200296SA, VU222520114SA, VU223120014SA, VU223120041SA, VU223420119SA, VU224050158SA, VU224070091SA, VU224200028SA, VU224420159SA, VU224520140SA, VU225170260SA, VU225420050SA, VU225520037SA, VU226040025SA, VU419320042SA, VU421490135SA, VU421490151SA, VU421490153SA, VU421490155SA, VU421490159SA, VU422460150SA, VU422460194SA, VU422460203SA, VU422460303SA, VU422480164SA, VU422500239SA, VU422500240SA, VU422500241SA, VU422500242SA, VU422500244SA, VU422500245SA, VU422500246SA, VU422500247SA, VU422500248SA, VU422500249SA, VU422500250SA, VU422500251SA, VU422500254SA, VU422500260SA, VU422500261SA, VU422500264SA, VU422500265SA, VU422500266SA, VU422500267SA, VU422500269SA, VU422510038SA, VU422510049SA, VU422510060SA, VU422510061SA, VU423050080SA, VU423050178SA, VU423220064SA, VU423220225SA, VU425130020SA, VU425130022SA, VU425130024SA, VU425130084SA, VU425130085SA, VU421490060SA, VU421490063SA, VU421490065SA, VU421490068SA, VU421490069SA, VU421490097SA, VU421490122SA, VU422460206SA, VU422460209SA, VU422460215SA, VU422460218SA, VU422460219SA, VU422460302SA, VU422460304SA, VU422460269SA, VU422460273SA, VU422460274SA, VU422460278SA, VU422460282SA, VU422460284SA, VU422480071SA, VU422460306SA, VU422460285SA, VU422480070SA, VU422460287SA, VU422460301SA, VU422480066SA, VU422480078SA, VU422480083SA, VU422480084SA, VU422480085SA, VU422480086SA, VU422480087SA, VU422480088SA, VU422480099SA, VU422480110SA, VU422480111SA, VU422480120SA, VU422480121SA, VU422480125SA, VU422480127SA, VU422480128SA, VU422480129SA, VU425130087SA, VU224200105SA, VU222120088SA, VU223130071SA, VU223220135SA, VU223220174SA, VU219490019SA, VU226050047SA, VU225460222SA, VU225460237SA, VU225470012SA, VU225470018SA, VU225470022SA, VU225470023SA, VU225470024SA, VU225470030SA, VU225470034SA, VU225470035SA, VU225470038SA, VU221050014SA, VU223310039SA, VU223450002SA, VU223310012SA, VU222190155SA, VU225060039SA, VU225060040SA, VU225060046SA, VU225060038SA, VU226154224SA, VU225500096SA, VU225500106SA, VU225500109SA, VU225500019SA, VU225500031SA, VU225500076SA.

View the original FDA recall notice

Why was this product recalled?

Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.

What actions did Ge Medical Systems Information Technologies, Inc. take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 6/29/2026 via US mail. The notice explained the issue, potential risk, and requested the following: 'Actions to be taken by Customer/User You can continue to use your device. Pending corrections from GE HealthCare, after pressing the Order Refresh icon do not rely on automatic closure of the Preview and or Review screen when: " patient has been disconnected, and " at least 2 minutes of system inactivity If this occurs, do the following to clear all patient information data: " On the Preview screen, select Reject > Start New Patient " On the Review screen, select Done > Start New Patient NOTE: Be certain Start New Patient is selected to clear all patient information data. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 30123 Digital Response Form or print, fill out manually, scan, and email to [email protected]." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology