Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2878-2026

GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system Recall

GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2878-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 10, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
11 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system

Affected model & UDI checker

Search the exact identifiers published for Z-2878-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5700592GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring systemCheck source noticeNot specified
Not specifiedGE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring systemCheck source notice00198953056752
FDA code information

GTIN 00198953056752, Lot/Serial Numbers: SWT25510001TA, 3M1D391215, 3M1D411081, 3M1D411083, 3M1D3B100D, SWT26110010TA, SWT26070001TA, SWT26120008TA, 3M1D411080, 3M1D411082, 3M1D3B100C.

View the original FDA recall notice

Why was this product recalled?

A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information (PHI).

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Ge Medical Systems Information Technologies, Inc. take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 7/10/2026 via traceable means. The notice explained the safety issue, and requested the following: "Actions to be taken by Customer/User You can continue to use Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.2.0, pending GE HealthCare correction. If you need help to identify if your Portrait Mobile Monitoring Solution has Portrait Core Services SW 1.2.0, contact GE HealthCare Service. Ensure that your facility follows all hospital networking and cybersecurity policies and procedures along with the security considerations outlined in the GE HealthCare Portrait Mobile Monitoring Solution Privacy and Security Manual p/n 5970026. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or a local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology