FDA lists this Class III recall for Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2880-2026
- Recall classification
- Class III
- Enforcement status
- Ongoing
- Initiation date
- Jun 26, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 8 US; 44 OUS
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL
Affected model & UDI checker
Search the exact identifiers published for Z-2880-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 6N319UL | Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL | Check source notice | Not specified |
| Not specified | Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL | Check source notice | 00842768001406 |
FDA code information
UDI-DI 00842768001406 Lot U1
Why was this product recalled?
Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
What actions did Siemens Healthcare Diagnostics, Inc. take?
On 6/26/2026, Siemens issued Urgent Medical Device Correction letters to all affected customers informing them of the issue and providing instructions. Customers are instructed to refer to the SDS which were created and provided with the Urgent Medical Device Corrections and the Urgent Field Safety Notices specifically for Syva Emit Disopyramide Assay Lot U1 for information regarding handling and disposal.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2880-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology