Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL Recall

Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

FDA recall Class III

Ongoing

FDA lists this Class III recall for Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2881-2026
Recall classification
Class III
Enforcement status
Ongoing
Initiation date
Jun 26, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
494 US; 114 OUS
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

Affected model & UDI checker

Search the exact identifiers published for Z-2881-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
7C219ULSyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULCheck source noticeNot specified
Not specifiedSyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULCheck source notice00842768014321
FDA code information

UDI-DI 00842768014321 Lot U2

View the original FDA recall notice

Why was this product recalled?

out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.

What actions did Siemens Healthcare Diagnostics, Inc. take?

On June 26, 2026, the firm began notifying customers via Urgent Medical Device Correction letters. Customers were informed of the issue and provided with revised Safety Data Sheets (SDS) specific to the impacted lots for handling and disposal information.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology