HardyDisk AST Cefiderocol 30g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii complex and Stenotrophomonas maltophilia to Cefiderocol Recall
HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30¿g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii complex and Stenotrophomonas maltophilia to Cefiderocol,
FDA recall Class II
Ongoing
FDA lists this Class II recall for HardyDisk AST Cefiderocol 30g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii complex and Stenotrophomonas maltophilia to Cefiderocol. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2887-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 8, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 43 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
HardyDisk AST Cefiderocol 30g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii complex and Stenotrophomonas maltophilia to Cefiderocol,
Affected model & UDI checker
Search the exact identifiers published for Z-2887-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Z9435 | HardyDisk AST Cefiderocol 30g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii complex and Stenotrophomonas maltophilia to Cefiderocol | Check source notice | Not specified |
FDA code information
Catalog Number: Z9435 UD-DI code: Pending Lot Number: 543461
Why was this product recalled?
Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Hardy Diagnostics take?
On 06/08/2026 an "URGENT: Medical Device Recall Notification" was sent via First Class Mail and Email and/or Facsimile to customer informing them that as a result of receiving a customer complaint/report, it was confirmed that antibiotic disk cartridges labeled as HardyDisk AST Cefiderocol (disk code FDC30) were found to contain sporadic disks labeled with Piperacillin/Tazobactam disk code (TZP110). Customers are instructed to: 1. Discard any remaining stock. 2. Response to this notification is requested within 5 business days of receipt of this letter, either through mail or email. To respond by mail, complete, sign, and send the enclosed inventory reconciliation form (FORM-2023C) via USPS First Class Mail. To respond by email, scan and send the completed inventory reconciliation form (FORM-2023C) to [email protected]. 3. Replacements will be sent out at no-charge to your laboratory or a credit will be applied to invoice upon request. For Questions - contact Technical Services Department at (800) 266-2222, option 2 or via email at [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2887-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology