5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905 Recall

5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905

FDA recall Class II

Ongoing

FDA lists this Class II recall for 5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2894-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 1, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905

Affected model & UDI checker

Search the exact identifiers published for Z-2894-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5D-459055-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905Check source noticeNot specified
Not specified5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905Check source notice05070002420378
FDA code information

UDI-DI: 05070002420378, Lot Number: KE132603

View the original FDA recall notice

Why was this product recalled?

Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.

What actions did Aniara Diagnostica LLC take?

Email notifications were sent to customers on 6/17/26. Please reply to this email as acknowledgment you have received this initial notification and please include the quantity of each product which is remaining in your possession. A notification letter dated 7/1/26 was sent to customers. Actions to be implemented: " Destroy all stock of 5-PLASMA Bivalirudin Control, 5-PLASMA Bivalirudin Calibrator, & 5-PLASMA AMR Verification Kits - Please permanently deface the box labels and vial labels before disposal. " Complete and return the Awareness Acknowledgement form. " If product has been consumed please add brief statement to the Comment section, complete the End User Information, and return the Awareness Acknowledgement form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology