Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Recall
Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Model/Catalog Number: 9323-5312 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2898-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 19, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 3
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Model/Catalog Number: 9323-5312 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A
Affected model & UDI checker
Search the exact identifiers published for Z-2898-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 9323-5312 | Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm | Check source notice | Not specified |
| Not specified | Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm | Check source notice | 04719886908077 |
FDA code information
Lot Code: Product number 9323-5312; UDI-DI 04719886908077; Lot number 25FC64
Why was this product recalled?
Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did UNITED ORTHOPEDIC CORP. take?
On 03/09/2026, the firm distributed an "URGENT: MEDICAL DEVICE RECALL" Letter informing customer that due the fact that the Femoral offset bushing in the affected devices is engraved with the incorrect numbering, and the rotational direction does not align with the intended orientation. Customers are instructed to: 1. Examine inventory immediately to locate the affected product whose quantities and lot numbers are listed in the letter. Note the product may have already been returned. 2. Cease all use of these devices immediately and quarantine all affected devices in their territory for immediate return. 3. Customer Service will be contacting customers regarding replacements. If customers still have the product, they are requested to immediately return the affected product using the prepaid airbills on RMA XXXX . a. Write the RMA number on the outside of the box clearly and legibly. b. Ship recalled product via Fed Ex standard overnight to: UOC USA INC. Attention: RMA# XXXX 16601 Hale Ave. Irvine, CA 92606 (Attached Shipping Label) 4. Additionally, if customers has further distributed the recall product, please provide that customer with a copy of this recall notification and assist them with the return of the product. 5. Complete and return the following page according to the instructions. For questions - contact UOC Sr. Director of Sales Operations at 949.328.3366.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2898-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology