Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2909-2026

Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LEF Recall

Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LEF

FDA recall Class II

Ongoing

FDA lists this Class II recall for Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LEF. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2909-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 17, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
12 units
Distribution
See key details
FDA cause
Labeling mix-ups

What is being recalled?

Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LEF

Affected model & UDI checker

Search the exact identifiers published for Z-2909-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1610-LEFBlom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LEFCheck source noticeNot specified
Not specifiedBlom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LEFCheck source notice00813307014133
FDA code information

Lot #0000369233/UDI: 00813307014133

View the original FDA recall notice

Why was this product recalled?

Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.

What actions did Freudenberg Medical, LLC take?

On June 26, 2026, Inhealth Technologies issued a Urgent Medical Device Recall Notification via FedEx to affected consignees. Inhealth ask consignees to take the following actions: 1. Do not use, ship, or distribute affected products 2. Identify and isolate all affected inventory 3. Destroy all affected devices 4. Complete the attached response form 5. Return confirmation no later than: July 8, 2026 6. Email confirmation of product destruction to: [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology