Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 Recall
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40
FDA recall Class II
Ongoing
FDA lists this Class II recall for Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2911-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 29, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 5069
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40
Affected model & UDI checker
Search the exact identifiers published for Z-2911-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| NG-0630-140-2 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
| NG-0640-140-2 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
| NG-0730-140-2 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
| NG-0740-140-2 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
| NG-NV-6-30 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
| NG-NV-6-40 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
| NG-NV-7-30 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
| NG-NV-7-40 | Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40 | Check source notice | Not specified |
FDA code information
REF/UDI-DI/Lot: NG-0630-140-2/00851616007583/Z2460578D,Z2460740B, Z2460872C, Z2460909F, Z2460948D, Z2460957Q; NG-0640-140-2/00851616007590/Z2460578E, Z2460647S, Z2460706A, Z2460746C, Z2460774A, Z2460865C, Z2460872B, Z2460901B, Z2460909D, Z2460909K, Z2460945H, Z2460948I, Z2460963N; NG-0730-140-2/00851616007606/Z2460578F, Z2460662A, Z2460740A , Z2460774F, Z2460850A, Z2460900B, Z2460945F, Z2460957R; NG-0740-140-2/00851616007613/Z2460578G, Z2460635C, Z2460656A, Z2460707D, Z2460733A, Z2460766A, Z2460774B, Z2460780A , Z2460865A, Z2460871A, Z2460872D, Z2460900D, Z2460909B, Z2460909I, Z2460948F, Z2460963T; NG-NV-6-30/00851616007644/Z2460776A, Z2460893A, Z2460909G, Z2460948E; NG-NV-6-40/00851616007651/Z2460776B, Z2460865D, Z2460865E, Z2460901C, Z2460909E, Z2460909J, Z2460957P, Z2460963K, Z2460963P; NG-NV-7-30/00851616007637/Z2460776C, Z2460789H, Z2460850B, Z2460900C, Z2460916A, Z2460948H, Z2460998A; NG-NV-7-40/00851616007620/Z2460776D, Z2460878A, Z2460872A, Z2460872E, Z2460901A, Z2460909C, Z2460909H, Z2460945G, Z2460948G, Z2460966B, Z2460963J;
Why was this product recalled?
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.
What actions did Contego Medical, Inc. take?
On 6/29/2026, recall notices started being delivered to healthcare professionals and risk managers by Medtronic sales representatives who informed of the following: 1) Sales representatives will visit, conduct inventory counts, quarantine, and remove unused affected devices. 2) If replacements are desired, they can be provided. 3) Complete and return the acknowledgment form via email to [email protected] Questions can be directed to the firm at [email protected] and 919-459-7250 Extension 4.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2911-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology