GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. Recall
GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.
FDA recall Class II
Ongoing
FDA lists this Class II recall for GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2915-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 22, 2026
- FDA report date
- Aug 19, 2026
- Quantity in commerce
- 13 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.
Affected model & UDI checker
Search the exact identifiers published for Z-2915-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| VSWVN1006 | GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. | Check source notice | Not specified |
FDA code information
Incorrect Catalogue Number/UDI: VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796; Lot Serial No. 30970948, 30970949, 30970950, 30970951, 30970952, 30970953, 30970954, 30970955, 30970956, 30970957, 30970958, 30970959, 30970960.
Why was this product recalled?
10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).
What actions did W. L. Gore & Associates, Inc. take?
On 05/22/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was sent via email to customers informing them that, 10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010). Customers are instructed to: . Identify and return any unused devices within the scope of this recall for replacement. . Complete and sign the enclosed CUSTOMER RESPONSE / DEVICE RECONCILIATION FORM and return to [email protected] within 2 weeks of receipt of this notification. . This notice needs to be shared with those who should be aware within your institution or to any organization where potentially affected devices have been transferred (as appropriate). . Implanting Physicians: If you have implanted one of the potentially mislabeled GORE VIABIL Short Wire Biliary Endoprosthesis devices, please be mindful of the potential health outcomes and plan to take the needed measures that could minimize any potential health hazards associated with this mislabeling. If a listed device has been used, please indicate the used device(s) on the CUSTOMER RESPONSE / DEVICE RECONCILIATION FORM and return to [email protected] For questions or further assistance - contact Julie Skaife, Global Product Specialist & Brand Manager at [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2915-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology